Diabetes
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow Request Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Velho, G.
Right arrow Articles by Reach, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Velho, G.
Right arrow Articles by Reach, G.
Social Bookmarking
 Add to CiteULike   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?

Diabetes, Vol 38, Issue 2 164-171, Copyright © 1989 by American Diabetes Association


ARTICLES

In vitro and in vivo stability of electrode potentials in needle-type glucose sensors. Influence of needle material

G Velho, P Froguel, R Sternberg, DR Thevenot and G Reach
Service de Diabetologie, Hotel-Dieu, Paris, France.

Enzymatic glucose sensors are based on the amperometric detection of an oxidable species generated during the oxidation of glucose by glucose oxidase. This measurement usually requires a working electrode (anode), an auxiliary electrode (cathode), and a reference electrode, the function of the latter being to keep constant the working potential of the anode, which is responsible for current generation. However, in the needle-type glucose sensors proposed so far, the reference electrode is missing, and its function is performed by the auxiliary electrode. We investigated, in vitro and in vivo in rats, the ability of several cathode-needle materials to behave as a reference electrode in two-electrode glucose sensors, i.e., to present a stable auxiliary electrode potential. In vitro, when glucose concentration was raised from 0 to 30 mM, the auxiliary potential of both gold- and silver-coated sensors presented a cathodic drift, whereas that of silver/silver chloride-coated sensors remained stable. In vivo, during insulin-induced hypoglycemia (5.9-2.4 mM), the auxiliary potentials of all sensors remained stable, whereas during glucose infusion (mean blood glucose concentration 11.2 mM), the auxiliary potentials of both gold- and silver-coated sensors presented an anodic drift, whereas those of silver/silver chloride-coated sensors remained stable. We also indirectly quantified the changes in sensor response induced by variations in the working potential in vitro and in vivo, simulating those that might be produced by a drift in the auxiliary potential. Such changes in the working potential could bring about a 30% unspecific variation in sensor response. We conclude that improvements in sensor analytical characteristics should be obtained with silver/silver-chloride-coated cathodes.