Diabetes 52:2790-2794, 2003
© 2003 by the American Diabetes Association, Inc.
Timing of Changes in Interstitial and Venous Blood Glucose Measured With a Continuous Subcutaneous Glucose Sensor
Michael S. Boyne,
David M. Silver,
Joy Kaplan, and
Christopher D. Saudek
From the Division of Endocrinology and Metabolism, Johns Hopkins University School of Medicine and the Applied Physics Laboratory, Johns Hopkins University, Baltimore, Maryland
The objective of this study was to use a subcutaneous continuous glucose sensor to determine time differences in the dynamics of blood glucose and interstitial glucose. A total of 14 patients with type 1 diabetes each had two sensors (Medtronic/MiniMed CGMS) placed subcutaneously in the abdomen, acquiring data every 5 min. Blood glucose was sampled every 5 min for 8 h, and two liquid meals were given. A smoothing algorithm was applied to the blood glucose and interstitial glucose curves. The first derivatives of the glucose traces defined and quantified the timing of rises, peaks, falls, and nadirs. Altogether, 24 datasets were used for the analysis of time differences between interstitial and blood glucose and between sensors in each patient. Time differences between blood and interstitial glucose ranged from 4 to 10 min, with the interstitial glucose lagging behind blood glucose in 81% of cases (95% CIs 72.5 and 89.5%). The mean (±SD) difference between the two sensors in each patient was 6.7 ± 5.1 min, representing random variation in sensor response. In conclusion, there is a time lag of interstitial glucose behind blood glucose, regardless of whether glycemia is rising or falling, but intersensor variability is considerable in this sensor system. Comparisons of interstitial and blood glucose kinetics must take statistical account of variability between sensors.
Address correspondence and reprint requests to Christopher D. Saudek, MD, Johns Hopkins University School of Medicine, Osler 576, 600 North Wolfe St., Baltimore, MD 21287. E-mail: csaudek{at}jhu.edu

CiteULike Del.icio.us Digg Reddit Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
B. Buckingham, D. M. Wilson, T. Lecher, R. Hanas, K. Kaiserman, and F. Cameron
Duration of Nocturnal Hypoglycemia Before Seizures
Diabetes Care,
November 1, 2008;
31(11):
2110 - 2112.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. R Lieberman, C. M Caruso, P. J Niro, G. E Adam, M. D Kellogg, B. C Nindl, and F M. Kramer
A double-blind, placebo-controlled test of 2 d of calorie deprivation: effects on cognition, activity, sleep, and interstitial glucose concentrations
Am. J. Clinical Nutrition,
September 1, 2008;
88(3):
667 - 676.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. Hermanides, I. M. Wentholt, A. A. Hart, J. B. Hoekstra, and J. H. DeVries
No Apparent Local Effect of Insulin on Microdialysis Continuous Glucose- Monitoring Measurements
Diabetes Care,
June 1, 2008;
31(6):
1120 - 1122.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. Kovatchev, S. Anderson, L. Heinemann, and W. Clarke
Comparison of the Numerical and Clinical Accuracy of Four Continuous Glucose Monitors
Diabetes Care,
June 1, 2008;
31(6):
1160 - 1164.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
I. M. Wentholt, J. B. Hoekstra, and J. H. DeVries
A Critical Appraisal of the Continuous Glucose-Error Grid Analysis
Diabetes Care,
August 1, 2006;
29(8):
1805 - 1811.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L. Allport, T. Baird, K. Butcher, L. MacGregor, J. Prosser, P. Colman, and S. Davis
Frequency and Temporal Profile of Poststroke Hyperglycemia Using Continuous Glucose Monitoring
Diabetes Care,
August 1, 2006;
29(8):
1839 - 1844.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
I. M. Wentholt, M. A. Vollebregt, A. A. Hart, J. B. Hoekstra, and J. H. DeVries
Comparison of a Needle-Type and a Microdialysis Continuous Glucose Monitor in Type 1 Diabetic Patients
Diabetes Care,
December 1, 2005;
28(12):
2871 - 2876.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
The Diabetes Research in Children Network (DirecNe
Eight-Point Glucose Testing Versus the Continuous Glucose Monitoring System in Evaluation of Glycemic Control in Type 1 Diabetes
J. Clin. Endocrinol. Metab.,
June 1, 2005;
90(6):
3387 - 3391.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. K. Nielsen, C. B. Djurhuus, C. H. Gravholt, A. C. Carus, J. Granild-Jensen, H. Orskov, and J. S. Christiansen
Continuous Glucose Monitoring in Interstitial Subcutaneous Adipose Tissue and Skeletal Muscle Reflects Excursions in Cerebral Cortex
Diabetes,
June 1, 2005;
54(6):
1635 - 1639.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. G. Pittas, R. Hariharan, P. C. Stark, C. L. Hajduk, A. S. Greenberg, and S. B. Roberts
Interstitial Glucose Level Is a Significant Predictor of Energy Intake in Free-Living Women with Healthy Body Weight
J. Nutr.,
May 1, 2005;
135(5):
1070 - 1074.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. P. Kovatchev, L. A. Gonder-Frederick, D. J. Cox, and W. L. Clarke
Evaluating the Accuracy of Continuous Glucose-Monitoring Sensors: Continuous glucose-error grid analysis illustrated by TheraSense Freestyle Navigator data
Diabetes Care,
August 1, 2004;
27(8):
1922 - 1928.
[Abstract]
[Full Text]
[PDF]
|
 |
|
Copyright © 2003 by the American Diabetes Association.
|
|
| |
|